CELL TESTER

Solar Cell Tester/DKSCT Series

  • Based on "IEC60904-1/JIS C 8913"
  • Based on "IEC60904-9/JIS C 8912"
  • Special characteristic measuring device for crystalline silicon photovoltaic cells
  • Applicable for R&D and Mass Production Lines

Solar Module Tester/DKSMT SeriesSolar Module EL Tester

Special characteristic inspection device for crystalline silicon photovoltaic cells

Standard configuration

  • Automatic mass production mode
  • Manual measurement mode
  • Automatic illuminance calibration function
  • Repeatability measurement function
  • Sweep measurement function
  • Classification setting
  • Parameter settings and data storage
  • Self-diagnostic function of I-V tester

Standard accessories

  • Solar simulator Class A
  • I-V Tester,PC(Windows7)
  • Monitor cell with temperature control function
  • Non-contact type radiation thermometer
  • Excel

Specifications

Input1φ AC200V to 220V, 50/60Hz 1000VA
I-V TesterOutput voltageOutput range3V Range:-3V to +3V
16V Range:-16Vto +7V
Output current-15A to +15A
Setting resolution1/30000 of Range
Setting accuracy+/-(0.05% FS+0.05% Setting)
Measurement voltageMeasurement range1V Range:-1V to +1V
2V Range:-2V to +2V
4V Range:-4V to +3V
8V Range:-8V to +3V
16V Range:-16V to +3V
Measurement resolution1/16000 of Range
Measurement accuracy+/-(0.05% FS+0.05% Setting)
Current measurementMeasurement range15A,7.5A,3A,1.5A,150mA,75mA,30mA,15mA
Measurement resolution1/15000 of Range
Measurement accuracy+/-(0.1% FS+0.1% Setting)
Measurement functionMeasurement modelight measurement: [Direction] Isc→Voc(positive),Voc→Isc(negative)
Dark measurement:[Direction] Isc→Voc, 0V →xV
Measurement parametersIsc, Voc, Pm, Vpm, Ipm, FF, η, Iv, Vi, Rs and Rsh
PCOSWindows 7 Professional
Dimensions
W*D*H
I-V Tester432mm*450mm*222mm Not including protrusion
PC425mm*490mm*175mm Not including protrusion
Weight of I-V Tester16Kg

Solar simulator Specifications

Spectral match0.75 to 1.25 Class A
Temporal instability of irradiance+/-1% Class A
Non-uniformity of irradiance+/-2% Class A
Irradiation Range180mm x 180mm or 200mm x 200mm
Lamp lifeAverage 1000 hours

Spectral match

Temporal instability of irradiance