DENKEN

SOLAR

Solar simulator for PV module / DKSMT Series

Photovoltaic Module Tester / DKSMT Series

Output measurement method for crystalline modules accordance with IEC60891/JIS C 8914 standard
Solar simulator accordance with IEC60904-9/JIS C 8912 standards
Output characteristic measurement system for crystalline solar cell modules
Available in R&D and mass production lines
Irradiation range: 1,200mm x 2,200mm applicable.

Solar simulator for PV module( Inspection equipment for crystalline solar cell module characteristics )

Solar simulator for PV module( Inspection equipment for crystalline solar cell module characteristics )
Standard structure
  • Automatic mass production mode
  • Manual measurement mode
  • Calibration for automatic irradiation.
  • Repeatability measurement function
  • Registration of multiple varieties
  • Rank Settings
  • Set each parameter and save data
  • Self-check function of I-V tester
Accessories
  • Class AAA solar simulator
  • I-V tester and PC included OS(Windows 10)
  • Monitor cell (with temperature control function)
  • Temperature measurement unit
  • Includes UPS

Specifications of I-V tester

Power source
1φ AC200V to 220V, 50/60Hz 3000VA or less
I-V Tester

Voltage output

Voltage output range 30V range:-0.3V to +30V
300V range: -0.3V to +300V
Current output range -0.3 A to +20 A
Maximum output capacity 400W
Setting resolution 1/30000 of range
Setting accuracy ±(0.1% of range + 0.1% of setting value)

Voltage measurements

Measurement range 3V range: -0.3V to +3V
7.5 V range: -0.3 V to +7.5 V
15V range: -0.3V to +15V
30V range: -0.3V to +30V
75V range: -0.3V to +75V
150V range: -0.3V to +150V
300V range: -0.3V to +300V
Measurement resolution 1/30000 of range
Measurement accuracy ±(0.1% of range + 0.1% of measured value)

Current measurement

Measuring range 20 A, 10 A, 5 A, 2.5 A, 200 mA
Measurement resolution 1/20000 of range
Measurement accuracy ±(0.1% of range + 0.1% of measured value)
irradiance measurement
Measurement range ±200mA
Measurement resolution ±0.02mA
Measurement accuracy ±(0.2mA + 0.1% of measured value)
Temperature measurement
Measurement points 3 measurement points (module, monitor cell and room temperature)
Others
OS Windows 10 Professional
Dimensions W x D x H (mm)
W×D×H (mm)
580 mm x 1700 mm x 700 mm not including protrusions

Specifications of the solar simulator

         
Spectral match (Class A) AM1.5G within ±25%
Temporal Instability(Class A)within ±1%
Non-uniformity of irradiance(Class A)within ±2%
Effective irradiation area up to~1100mm×2000mm
Pulse light irradiation time : up to10ms or up to100ms *Depends on the model.
Lamp lifeaverages 100,000 flashes
Measurement interval Minimum 15 seconds (90 seconds when initially charged)

Spectral match

Spectral match