DENKEN

SOLAR

Solar simulator for PV cell / DKSCT Series

Solar Cell Tester / DKSCT Series

Output measurement method for crystalline cells based on IEC60904-1/JIS C 8913 standards
Photovoltaic solar simulator based on IEC60904-9/JIS C 8912 standards
Output characteristic tester for crystalline solar cells
Available in R&D and mass production lines

Solar simulator for PV cell (Crystalline Solar Cell Characteristic Inspection Equipment)

Solar simulator for PV cell (Crystalline Solar Cell Characteristic Inspection Equipment)
Standard configurations
  • Automatic mass production mode
  • Manual measurement mode
  • Automatic illuminance calibration function
  • Repeatability measurement function
  • Sweep measurement
  • Rank settings
  • Set each parameter and save data
  • Self-check function of I-V tester
Accessories
  • Class AAA solar simulator
  • I-V tester and PC included OS(Windows 10)
  • Monitor cell (with temperature control function)
  • Non-contact radiation thermometer
  • Excel

Specifications of I-V tester

Power source
1φ AC200V ~ 220V, 50/60Hz 1000VA
I-V Tester

Output of voltage

Voltage output range 3V range: -3V to +3V / 16V range: -16V to +3V
Current output range -15A to +15A
Setting resolution 1/30000 of range
Setting accuracy ±(0.05% of range + 0.05% of setting value)

Voltage measurement

Measurement range 1V range: -1V to +1V
2V range: -2 V to +2 V
4V range: -4V to +3V
8V range: -8V to +3V
16V range: -16V to +3V
Measurement resolution 1/16000 of range
Measurement accuracy ±(0.05% of range + 0.05% of measured value)

Current measurement

Measuring range 15A, 7.5A, 3A, 1.5A, 150mA, 75mA, 30mA and 15mA
Measurement resolution 1/15000 of range
Measurement accuracy ±(0.1% of range + 0.1% of measured value)
Measurement function
Measurement with light : Isc→Voc forward, Voc→Isc reverse
Dark measurement : Isc→Voc direction, 0V→ -xV direction
Measured items Isc, Voc, Pm, Vpm, Ipm, FF, η, Iv, Vi, Rs and Rsh
Others
OS Windows 10 Professional
Dimensions W x D x H (mm)
I-V tester432mm x 450mm x 222mm not including protrusions
PC425mm x 490mm x 175mm not including protrusions
Weight (I-V tester)
16Kg

Specifications of the solar simulator

         
Spectral match (Class A)AM1.5G Within 0.75 to 1.25
Instability of irradiance (Class A)within ±1%
Non-uniformity of irradiance (Class A)within ±2%
Effective irradiation area180mm*180mm or 200mm*200mm
Lamp life: Average 1000 hours

Spectral match

Spectral match

Instability of irradiance

Instability of irradiance